Artificial intelligence as a second reader for screening mammography

E Nakai, Y Miyagi, K Suzuki… - Radiology …, 2024 - academic.oup.com
E Nakai, Y Miyagi, K Suzuki, A Scoccia Pappagallo, H Kayama, T Matsuba, L Yang, S Xu
Radiology Advances, 2024academic.oup.com
Background Artificial intelligence (AI) has shown promise in mammography interpretation,
and its use as a second reader in breast cancer screening may reduce the burden on health
care systems. Purpose To evaluate the performance differences between routine double
read and an AI as a second reader workflow (AISR), where the second reader is replaced
with AI. Materials and Methods A cohort of patients undergoing routine breast cancer
screening at a single center with mammography was retrospectively collected between 2005 …
Background
Artificial intelligence (AI) has shown promise in mammography interpretation, and its use as a second reader in breast cancer screening may reduce the burden on health care systems.
Purpose
To evaluate the performance differences between routine double read and an AI as a second reader workflow (AISR), where the second reader is replaced with AI.
Materials and Methods
A cohort of patients undergoing routine breast cancer screening at a single center with mammography was retrospectively collected between 2005 and 2021. A model developed on US and UK data was fine-tuned on Japanese data. We subsequently performed a reader study with 10 qualified readers with varied experience (5 reader pairs), comparing routine double read to an AISR workflow.
Results
A “test set” of 4,059 women (mean age, 56 ± 14 years; 157 positive, 3,902 negative) was collected, with 278 (mean age 55 ± 13 years; 90 positive, 188 negative) evaluated for the reader study. We demonstrate an area under the curve =.84 (95% confidence interval [CI], 0.805-0.881) on the test set, with no significant difference to decisions made in clinical practice (P = .32). Compared with routine double reading, in the AISR arm, sensitivity improved by 7.6% (95% CI, 3.80-11.4; P = .00004) and specificity decreased 3.4% (1.42-5.43; P = .0016), with 71% (212/298) of scans no longer requiring input from a second reader. Variation in recall decision between reader pairs improved from a Cohen kappa of κ = .65 (96% CI, 0.61-0.68) to κ = .74 (96% CI, 0.71-0.77) in the AISR arm.
Conclusion
AISR improves sensitivity, reduces variability and decreases workload compared to routine dual screening.